Comparison of IOL Materials

From EyeWiki

All content on Eyewiki is protected by copyright law and the Terms of Service. This content may not be reproduced, copied, or put into any artificial intelligence program, including large language and generative AI models, without permission from the Academy.


Introduction

Intraocular lenses (IOLs) are commonly implanted after surgical removal of a cataractous lens. Since the introduction of the first modern IOL in 1949, IOL materials and designs have evolved to encompass medical needs. A variety of IOL materials are currently available, including collamer, hydrophobic acrylic, hydrophilic acrylic, PEG-PEA/HEMA/Styrene copolymer, polymethylmethacrylate (PMMA), and silicone . While PMMA was the first material to be used in IOL implantation, technological advances have brought about silicone and acrylic lenses, which are foldable and therefore, more applicable with the advent of small-incision, phacoemulsification-assisted cataract surgery. These IOL materials contain distinct physical and optical properties, with unique advantages and disadvantages.

IOL Properties

IOL materials can be classified using various properties they possess including abbe number, biocompatibility, hydrophobicity, hygroscopy, glass transition temperature, and refractive index. These properties contribute to visual quality.

Abbe Number

The Abbe number is a measure of an optical material's chromatic dispersion, with higher values representing lower chromatic dispersion, and better optical quality [1]. As the refractive index increases, the Abbe number decreases [2]. The Abbe number of the natural crystalline lens is 47 [1]. The human visual system exhibits spherical refractive errors of 2.5D within the visible spectrum; however, the phakic human eye is well adapted to this longitudinal chromatic aberration (LCA) due to compensatory mechanisms, including blue-light filtering capabilities of the lens and macula [3][4]. Moreover, chromatic dispersion does not vary significantly from one person's eye to another [5]. In pseduophakic eyes, however, various IOL materials determine LCA, with Abbe numbers ranging from 37-55 [5]. Accommodation, contrast sensitivity, emmetropization, and visual acuity are all affected by chromatic aberration [4].

Biocompatibility

Biocompatibility is defined as the capability of an implanted IOL material to remain optically functional in vivo without causing significant adverse responses such as anaphylaxis, inflammation, necrosis, rejection, thrombosis, or tumorigenesis [6]. IOL materials must also block ultraviolet rays, be optically transparent, and have a high refractive index[7]. Capsular biocompatibility is determined by direct contact of the IOL with remnant lens epithelial cells (LECs), which may lead to adherence and anterior capsule opacification (ACO) and posterior capsule opacification (PCO)[8]. Uveal biocompatibility refers to how well an IOL is tolerated in the eye without causing immunogenic responses of the anterior choroid, ciliary body, and iris [8].

Hydrophobicity

An IOL material surface may be hydrophilic or hydrophobic, measured by photographing a water droplet placed onto the material surface. The contact angle measures the degrees between the material’s surface and water droplet’s surface[7]. Large contact angles define hydrophobic materials, whereas smaller angles define hydrophilic materials[7].

Hygroscopy

Hygroscopy is the ability of a material to absorb and hold water (also known as a lens material's equilibrium water content) [7]. Increasing a material’s hygroscopy reduces the common postoperative appearance of glistenings, which are small fluid-filled microvaculoes that appear after IOL implantation. This is likely because once water enters the material, it interacts with hydrophilic groups and therefore does not collect in microvacuoles [9].

Glass transition temperature

The fluidity of polymers changes with temperature. A polymer appears solid at lower temperatures because its chains move slowly, whereas it transitions to a viscous material at higher temperatures due to increased energy and molecular mobility [7]. The glass transition temperature of a polymer is the temperature at which the polymer changes from a rigid, solid-like glassy fluid to a softer, rubber-like fluid [7]. IOLs are manufactured with a goal glass transition temperature below the physiologic body temperature of 37°C and normal room temperature because materials with a glass transition temperature above 37°C or normal room temperature would not unfold at physiologic body temperatures or at room temperature [6].

Refractive index

The refractive index is a measure of the bending of a ray of light, calculated by the ratio of the velocity of light within a vacuum to the velocity of light within the material [10]. The refractive index of the natural crystalline lens is 1.4 [1]. The refractive index increases with the addition of halogens, aromatic groups, or sulfur [2]. An inverse relationship exists between the refractive index and IOL thickness [10].

IOL Materials

Collamer

The lens material, Collamer® is a co-polymer containing collagen. Collamer® is a registered brand name proprietary to STAAR Surgical. More specifically, Collamer® is a hydroxyethyl methacrylate (HEMA)/porcine collagen containing biocompatible polymer material that is widely used as foldable intraocular lens (IOL) in the posterior chamber as a phakic IOL. These are referred to as implantable collamer lenses, with common brand names including EVO ICL, EVO Toric ICL, EVO+ ICL, EVO+ Toric ICL (with "+" designating a larger optical zone). The EVO family of ICLs contains a UV absorber made from a UV absorbing material; EVO ICL has been available outside the US since 2014 and granted FDA approval in the US in 2022. In the US, EVO/EVO+ spherical and toric ICLs, with the 0.36 mm diameter central port design, was introduced following FDA approval in 2022. This replaced the prior Visian ICL, which did not have the central hole design and required preoperative peripheral iridotomies.

The success of collamer IOL also depends upon the lens vault. Lens vault (the distance between the anterior surface of the crystalline lens and the posterior surface of the ICL lens) should be assessed 24 hours postoperatively at a slit lamp. Although the postoperative vault of the ICL lens is intended to be approximately equal to the central corneal thickness, it is generally accepted that the optimal vault should be between 50% and 150% of central corneal thickness, which is equivalent to a range of 250 to 900 microns. [11] [12] However, in the absence of symptoms, lens vault outside of this range may not necessarily require exchange or removal.

Theoretically, a low vault value may induce anterior subcapsular cataract formation due to mechanical contact between the IOL; conversely, a high vault value increases the risk of angle closure glaucoma secondary to chronic pigment dispersion from contact of the collamer IOL on the posterior surface of the iris [11] [12][13]. Patient age, anterior chamber depth, sulcus-to-sulcus distance, white-to-white distance, and IOL size are additional factors related to the vault [12]. There are no pupil size requirements for ICL implantation. Benefits of ICL compared to laser corneal refractive surgery, include ability to correct higher degrees of myopia, higher degrees of astigmatism, reducing very high myopia, and can be implanted in eyes with thinner corneas on pachymetry. Adverse effects with phakic IOL implanted in the posterior chamber include increased incident of anterior subcapsular cataract formation, uncorrected residual astigmatism, chronic uveitis, corneal endothelial cell loss, endophthalmitis, and pupil ovalization [13].

The FDA approval criteria for the EVO ICL family of lenses is indicated for use in phakic eyes in patients 21-45 years of age who meet the following criteria:

  • Myopic reduction (with or without astigmatism) with a spherical equivalent of -3.0 D to -20.0 D (and astigmatic correction of 1.0 D to 4.0 D at the spectacle plane; the cylinder power is up to 6D outside the US)
  • Anterior chamber depth (as measured from the corneal endothelium to the anterior lens capsule) ≥ 3.0 mm
  • Stable refractive error of ≤ 0.5 D within 1-year prior to implantation


The safety and effectiveness results from the US FDA EVO ICL clinical trial can be found here[14]. The directions for use for the EVO ICL can be found here.[15] .

Hydrophobic Acrylic

Introduced in 1993, hydrophobic acrylic IOLs such as AcrySof are the most widely used IOL material currently. They are composed of crosslinked copolymers of acrylic esters and other acrylic ester co-monomers, with a carbon backbone and ester side groups [6]. Single and three-piece lenses can be folded and implanted through small incisions while maintaining their original shape[7][16]. Long-term studies have demonstrated that patients implanted with hydrophobic acrylic lenses have lower rates of, and less dense, PCO, and were less likely to require Nd-YAG capsulotomy than patients implanted with hydrophilic acrylic IOLs [17]. This may be due to the tendency of hydrophobic acrylic IOLs to adhere to the posterior lens capsule through fibronectin bindings, leading to decreased space for LEC migration to occur in between the IOL and posterior capsule [17] [18]. Unfortunately, glistenings commonly affect hydrophobic acrylic IOLs [19].

Hydrophilic Acrylic

The new generation of foldable IOLs are composed of the methacrylate backbone of PMMA with additional hydroxyl groups introduced in the side chains [7]. The addition of hydroxyethylmethacrylate (HEMA), a material used in the manufacturing of contact lenses, poly(2-HEMA), or poly-HEMA confers flexibility to the IOL [7][10]. With the development of small-incision phacoemulsification assisted cataract surgery, foldable IOLs are the main type of lens used. Since these lenses are highly flexible, they require an incision of approximately 1.8mm, thereby making them highly beneficial in microincision cataract surgery [20].

Hydrogel lenses, as in those with poly-HEMA chains [10], have varying optical properties depending on their water content [7]. Once the manufactured lens, which is dry and opaque, is placed in water, it becomes soft and optically clear [7]. Hydrophilic IOLs have a high water content of 18-34%, leading to superior biocompatibility with lower rates of glare, a lower refractive index of 1.40-1.43 and an increased IOL thickness (Table 1) [7]. In a prospective study of 86 eyes with pseudoexfoliation syndrome (PEX) comparing hydrophilic acrylic, hydrophobic acrylic and silicone IOLs, hydrophilic acrylic IOLs had superior capsular biocompatibility since lens epithelial cell outgrowth was lowest with this group [21]. However, hydrophilic acrylic IOL had poor uveal biocompatibility since this group had the highest deposition of debris on the IOL surface after 12 to 18 months, and the highest rate of PCO [21]. This may be because the swelling that occurs within the lens after manufacturing induces difficulty maintaining a sharp posterior edge [7]. Although older generations of hydrophilic acrylic IOLs were associated with primary internal or surface calcifications due to calcium and phosphoric acid deposition, this does not seem to be a problem in newer generations [7][20].

PEG-PHS/HEMA/Styrene (PHS) Copolymer

The appeal of hydrophilic IOL materials incorporating HEMA derives from its biocompatibility, flexibility for microincisions, and lower rates of glare, but is set back by the increased incidence of PCO and calcification [22].  In contrast, the lower rates of PCO occurrence and surgical complications are more advantageous with hydrophobic acrylic IOL material yet are more likely to form glistenings in the long term [21] [22]. As lenses with low hygroscopy, 0.1% - 0.5%, are more likely to develop glistenings [9]. Companies have sought to maximize the balance in incorporating both hydrophilic and hydrophobic materials to further advance IOL material configuration. For instance, enVista’s MX60 IOL is composed of a copolymer mix of 40% hydrophobic poly(ethylene glycol)-phenyl ether acrylate (PEG-PEA), 30% of hydrophilic HEMA, 26%  of styrene to add bulk and increase the refractive index, and 4% of ethylene glycol dimethacrylate (EG-DMA) for structural integrity [7]. The main composition of enVista MX60L IOL, PEG-PEA, HEMA, and styrene (PHS Copolymer), is a hydrophobic IOL material with significant hydrophilic content that holds a hardness of 1.8MPa and a hygroscopy of 4% [7]. This is partly attributed to the PEG-PEA repeating subunit of ethylene glycol allowing a carbon chain of flexibility and hydrophobicity and HEMA’s terminal hydrophilic hydroxyl group increasing the material’s hygroscopy. Studies have shown that usage of enVista MX60L in cataract surgery demonstrates a good safety profile, stable refractive outcomes, low incidences of PCO and incidences of Nd:YAG capsulotomies, and no incidences of glistenings [23] [24].

PMMA

PMMA, a non-foldable, rigid, hydrophobic material, is available in one-piece version and three-piece versions that is lathe cut from PMMA buttons or rods [6]. PMMA, the first IOL material implanted into a human by Harold Ridley in 1949, is known for its excellent tissue tolerance, long-term stability, and reduced cost [16]. It has a contact angle of 65° to 71°, hygroscopy percentage of 0.4% to 0.8%, glass transition temperature of 105°C to 113° C, and a high refractive index of 1.49 (Table 1). Furthermore, PMMA IOLs have superior optical clarity, allowing for a broad spectrum of light to be transmitted [16].

Despite these properties, the PMMA IOL has been replaced by modern foldable lenses. PMMA's rigid structure prevents it from being folded to enter the small 2 mm incision made for phacoemulsification; a wider incision of approximately 5.5mm to 6mm is required for implantation, which resulting in delayed healing, poor intraocular pressures, and postoperative astigmatism [10]. A well-known long-term complication of PMMA IOLs is snowflake degeneration, which is an indication of PMMA IOL explantation[25]. PMMA's hydrophobicity may be damaging to the corneal endothelium secondary to adherence to these cells during implantation.[10] PMMA and hydrogel IOLs are associated with higher rates of PCO compared to acrylic and silicone IOLs [26][27]. While this may in part be due to IOL surface roughness providing a scaffold for PCO tissue formation, recent studies suggest that that IOL edges may play a more significant role; long-term PCO and neodymium-doped ytrium aluminum garnet (Nd:YAG) capsulotomy rates between square-edge PMMA and round-edge PMMA IOLs found that a squared posterior optic edge reduced proliferation of lens epithelial cell on the posterior lens capsule, thereby reducing PCO rates [26]. PMMA IOLs are currently used as scleral-sutured IOLs for instance when capsular implantation is not feasible [28].

Silicone

A synthetic polymer composed of a backbone of repeating silicon-oxygen groups, silicone is used to make foldable IOLs [10]. Silicone lenses are made with plate haptics or modified C-loop haptics. Although plate haptics have lower PC and cystoid macula edema rates, an intact anterior capsule is required to prevent decentering [10].

Silicone has the desirable properties of flexibility, biocompatibility, autoclavability, relative inertness, and chemical stability [29]. Silicone is hydrophobic, with a contact angle of 97° to 120°, hygroscopy percentage of 0.38%, and glass transition temperature of -120°C to -90°C (Table 1). Since silicone lenses have a refractive index of 1.43, which is lower than that of acrylic lenses, they are thicker for the same refractive power - this thickness may require an incision of up to 3.2 mm for implantation [10]. A review article by Kwon et al. demonstrated that silicone IOLs had lower PCO and Nd:YAG capsulotomy rates than hydrophobic acrylic IOLs at long-term use (>6 years) following implantation [30]. This may be because LECs are less likely to adhere to the silicone IOL surface, and because silicone IOLs have the squarest edges [7].

However, silicone is not widely used as an IOL biomaterial due to its susceptibility to adherence of bacteria, cells, and silicone oils [29][31]. This may lead to endophthalmitis, capsular opacification, IOL decentration, and capsular contraction, thereby impacting patients' post-operative visual acuity [31]. Silicone IOLs lead to premature, uncontrolled opening of the IOL within the anterior chamber, which can make IOL implantation more complex; furthermore, silicone IOLs may fog with condensation during subsequent vitrectomies and thus should be used in caution in diabetic eyes [6][16]. Deposition of calcium oxalate on the silicone IOL has been described in eyes with asteroid hyalosis, following YAG capsultomy [32].

Table 1

IOL material IOL Design Implanted Location Sphericity Status Hygroscopy (%) Glass Transition Temperature (°C) Refractive Index (n) Complication
Collamer 1-piece Sulcus Negative 40 40 1.44
Hydrophobic acrylic 1-piece Capsular bag Spheric 0.1-0.5 16 - 55 1.47-1.56 Glistenings
Hydrophilic acrylic 1-piece Capsular bag Neutral 18-38 10 - 20 1.40-1.43
PEG-PEA/HEMA/Styrene copolymer 1-piece Capsular bag Neutral 4-5 28 1.54
PMMA 1-piece Anterior chamber Negative 0.4-0.8 105 - 113 1.49 Snowflake degeneration[25]
Silicone 3-piece Capsular bag Neutral 0.38 -90 - -120 1.43 Deposition of calcium oxalate from asteroid hyalosis following YAG capsulotomy[32]
  1. 1.0 1.1 1.2 Zhao H, Mainster MA. The effect of chromatic dispersion on pseudophakic optical performance. British journal of ophthalmology. 2007;91:1225-1229.
  2. 2.0 2.1 Helmstetter S, Badur T, Hampp N. High-refractive quinolinone-based polymers for ophthalmic devices. Journal of polymer research. 2016;23:1.
  3. Millán MS, Vega F, Poyales F, Garzón N. Clinical assessment of chromatic aberration in phakic and pseudophakic eyes using a simple autorefractor. Biomedical optics express. 2019;10:4168-4178.
  4. 4.0 4.1 Eppig T, Rawer A, Hoffmann P, Langenbucher A, Schröder S. On the Chromatic Dispersion of Hydrophobic and Hydrophilic Intraocular Lenses. Optometry and vision science. 2020;97:305-313.
  5. 5.0 5.1 Bradley A, Xu R, Wang H, et al. The Impact of IOL Abbe Number on Polychromatic Image Quality of Pseudophakic Eyes. Clinical ophthalmology (Auckland, N.Z.). 2020;14:2271-2281.
  6. 6.0 6.1 6.2 6.3 6.4 Tripti D, Haldar RS, Geetha S, Niyogi UK, Khandal RK. Materials for intraocular lenses (IOLs): Review of developments to achieve biocompatibility. e-Polymers. 2009;9:1466-1488.
  7. 7.00 7.01 7.02 7.03 7.04 7.05 7.06 7.07 7.08 7.09 7.10 7.11 7.12 7.13 7.14 7.15 7.16 Tetz M, Jorgensen MR. New hydrophobic IOL materials and understanding the science of glistenings. Current eye research. 2015;40:969-981.
  8. 8.0 8.1 Tan X, Zhan J, Zhu Y, et al. Improvement of Uveal and Capsular Biocompatibility of Hydrophobic Acrylic Intraocular Lens by Surface Grafting with 2-Methacryloyloxyethyl Phosphorylcholine-Methacrylic Acid Copolymer. Scientific reports. 2017;7:40462-40462.
  9. 9.0 9.1 Yildirim TM, Fang H, Schickhardt SK, Wang Q, Merz PR, Auffarth GU. Glistening formation in a new hydrophobic acrylic intraocular lens. BMC ophthalmology. 2020;20:186-186.
  10. 10.0 10.1 10.2 10.3 10.4 10.5 10.6 10.7 10.8 Kapoor S and Gupta S. Basic science of intraocular lens material. In: Wang X, Ferreri F, eds. Intraocular Lens. IntechOpen. 2020.
  11. 11.0 11.1 Alfonso JF, Fernández-Vega L, Lisa C, Fernandes P, González-Meijome J, Montés-Micó R. Long-term evaluation of the central vault after phakic Collamer® lens (ICL) implantation using OCT. Graefe's archive for clinical and experimental ophthalmology. 2012;250:1807-1812.
  12. 12.0 12.1 12.2 Kim BK, Chung YT. Clinical results of Visian implantable collamer lens implantation according to various sizes and implantation angles. European journal of ophthalmology. 2022;2021;32:2041-2050.
  13. 13.0 13.1 Uozato H, Shimizu K, Kawamorita T, Ohmoto F. Modulation transfer function of intraocular collamer lens with a central artificial hole. Graefe's archive for clinical and experimental ophthalmology. 2011;249:1081-1085.
  14. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9738966/pdf/opth-16-3981.pdf
  15. file:///Users/jeremykudrna/Downloads/eDFU-0001_Rev_03_EVO-EVO+%20Visian%20ICL%20(1).pdf
  16. 16.0 16.1 16.2 16.3 Wang B, Lin Q, Shen C, Han Y, Tang J, Chen H. Synthesis of MA POSS-PMMA as an intraocular lens material with high light transmittance and good cytocompatibility. RSC advances. 2014;4:52959-52966.
  17. 17.0 17.1 Chang A, Kugelberg M. Posterior Capsule Opacification 9 Years after Phacoemulsification with a Hydrophobic and a Hydrophilic Intraocular Lens. European journal of ophthalmology. 2017;27:164-168.
  18. Li N, MD, Chen X, MD, Zhang J, MD, et al. Effect of AcrySof versus Silicone or Polymethyl Methacrylate Intraocular Lens on Posterior Capsule Opacification. Ophthalmology (Rochester, Minn.). 2008;115:830-838.
  19. Colin J, Orignac I, Touboul D. Glistenings in a large series of hydrophobic acrylic intraocular lenses. J Cataract Refract Surg. 2009 Dec;35(12):2121-6. doi: 10.1016/j.jcrs.2009.06.029. PMID: 19969218.
  20. 20.0 20.1 Bompastor-Ramos, Paula, MD, MSc, Póvoa J, MD, Lobo, Conceição, MD, PhD, et al. Late postoperative opacification of a hydrophilic–hydrophobic acrylic intraocular lens. Journal of cataract and refractive surgery. 2016;42:1324-1331.
  21. 21.0 21.1 21.2 Richter-Mueksch S, MD, Kahraman G, MD, Amon M, MD, Schild-Burggasser G, MD, Schauersberger J, MD, Abela-Formanek C, MD. Uveal and capsular biocompatibility after implantation of sharp-edged hydrophilic acrylic, hydrophobic acrylic, and silicone intraocular lenses in eyes with pseudoexfoliation syndrome. Journal of cataract and refractive surgery. 2007;33:1414-1418.
  22. 22.0 22.1 Luo C, Wang H, Chen X, Xu J, Yin H, Yao K. Recent Advances of Intraocular Lens Materials and Surface Modification in Cataract Surgery. Front Bioeng Biotechnol. 2022 Jun 8;10:913383. doi: 10.3389/fbioe.2022.913383. PMID: 35757812; PMCID: PMC9213654.
  23. Heiner P, Ligabue E, Fan A, Lam D. Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) - results of a European and Asian-Pacific study. Clin Ophthalmol. 2014 Mar 27;8:629-35. doi: 10.2147/OPTH.S56135. PMID: 24729678; PMCID: PMC3974695.
  24. Packer M, Fry L, Lavery KT, Lehmann R, McDonald J, Nichamin L, Bearie B, Hayashida J, Altmann GE, Khodai O. Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista). Clin Ophthalmol. 2013;7:1905-12. doi: 10.2147/OPTH.S50499. Epub 2013 Sep 24. PMID: 24109169; PMCID: PMC3792945.
  25. 25.0 25.1 Apple DJ, Peng Q, Arthur SN, Werner L, Merritt JH, Vargas LG, Hoddinott DS, Escobar-Gomez M, Schmidbauer JM. Snowflake degeneration of polymethyl methacrylate posterior chamber intraocular lens optic material: a newly described clinical condition caused by unexpected late opacification of polymethyl methacrylate. Ophthalmology. 2002 Sep;109(9):1666-75. doi: 10.1016/s0161-6420(02)01122-3. PMID: 12208715.
  26. 26.0 26.1 Haripriya A, Chang DF, Vijayakumar B, et al. Long-term Posterior Capsule Opacification Reduction with Square-Edge Polymethylmethacrylate Intraocular Lens: Randomized Controlled Study. Ophthalmology (Rochester, Minn.). 2017;124:295-302.
  27. Cheng JW, Wei RL, Cai JP, Xi GL, Zhu H, Li Y, Ma XY. Efficacy of different intraocular lens materials and optic edge designs in preventing posterior capsular opacification: a meta-analysis. Am J Ophthalmol. 2007 Mar;143(3):428-36. doi: 10.1016/j.ajo.2006.11.045. Epub 2006 Dec 28. PMID: 17224119.
  28. Lyu J, Zhao P. Simplified ab externo fixation technique to treat late dislocation of scleral-sutured polymethyl methacrylate intraocular lenses. Eye (London). 2016;30:668-672.
  29. 29.0 29.1 Huang X, Yao K, Zhang H, Huang X, Xu Z. Surface modification of silicone intraocular lens by 2-methacryloyloxyethyl phosphoryl-choline binding to reduce Staphylococcus epidermidis adherence. Clinical & experimental ophthalmology. 2007;35:462-467.
  30. Kwon YR, Hwang YN, Kim SM. Posterior Capsule Opacification after Cataract Surgery via Implantation with Hydrophobic Acrylic Lens Compared with Silicone Intraocular Lens: A Systematic Review and Meta-Analysis. Journal of ophthalmology. 2022;2022:3570399-3570399.
  31. 31.0 31.1 Yao K, Huang X, Huang X, Xu Z. Improvement of the surface biocompatibility of silicone intraocular lens by the plasma-induced tethering of phospholipid moieties. Journal of Biomedical Materials Research Part A. 2006;78A:684-692.
  32. 32.0 32.1 Fuchihata M, Saito Y, Kitaguchi Y, Nishida K, Hamanaka N, Sakamoto Y, Shiotani Y, Nakae K, Kiuchi Y, Tatebayashi M. [Calcification on the posterior surface of an implanted silicone intraocular lens in a patient with asteroid hyalosis]. Nippon Ganka Gakkai Zasshi. 2006 Sep;110(9):736-40. Japanese. PMID: 17025232.